EasyManuals Logo

Keithley DMM7510 Calibration Manual

Keithley DMM7510
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
Section
4: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
4-10 DMM7510-905-01 Rev. A / April 2015
cal.adjust.count
This attribute returns the number of times the instrument has been adjusted.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Nonvolatile memory Not applicable
Usage
adjustments = cal.adjust.count
adjustments
The number of adjustments
Details
You can use this command if calibration is locked or unlocked.
Example
Count = cal.adjust.count
print(Count)
Assign the number of times the instrument has been adjusted
to a user variable named Count.
Output the value. Example output:
3
This shows that the instrument has been adjusted 3 times.
Also see
cal.adjust.date (on page 4-11)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals