Section
2: Performance verification Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
2-36 DMM7510-905-01 Rev. A / April 2015
Verify the simulated RTD temperature
To verify RTD accuracy, use the following setup:
1. For 4-wire accuracy, connect the Model DMM7510 INPUT and SENSE terminals to the calibrator
as shown in the following figure.
2. For 3-wire accuracy, connect the Model DMM7510 INPUT and SENSE terminals to the calibrator
as shown in the following figure.
The SENSE HI wire is not required for 3-wire RTD measurements. For 3-wire RTD, accuracy is for
< 0.1 Ω lead resistance mismatch for input HI and LO. Add 0.25 °C per 0.1 Ω of HI-LO lead
resistance mismatch.