EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Calibration Manual

Keithley DMM7510
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Section
3: Calibration Model DMM7510 7½ Digit Graphical Sampling Multimeter
Calibration Manual
3-22 DMM7510-905-01 Rev. A / April 2015
--Rear DC adjustment step 3: DCI 10 A full-scale adjustment
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear() -- Clear the error event log and Home screen error triangle.
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step #3, 2A, DCA 10A range")
cal.adjust.rear.dc(3, 2.0) -- Requires calibrator to source 2A.
code,message,severity = eventlog.next() -- Set variables to print the error
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step #3, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')
--Rear AC adjustment step 1: ACI 10 A Zero adjustment
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear() -- Clear the error event log and Home Screen error triangle.
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step#1, 0.5A, ACI 10A Zero")
cal.adjust.rear.ac(1) -- Requires Fluke Model 5700 to source 0.5A ac at 1KHz into
10A and Lo rear terminals.
code,message,severity = eventlog.next() -- Set variables to print the error
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Rear, Step#1, ACI, Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals