2400 Series SourceMeter
®
User’s Manual Relative and Math 8-3
Remote rel programming
Rel commands
Table 8-1 summarizes rel commands. See Section 18, “CALCulate2” subsystem, for addi-
tional information.
Rel programming example
Table 8-2 lists commands for setting up and enabling rel. These commands set up the
SourceMeter as follows:
Rel value: 5
Rel state: enabled
Table 8-1
Rel commands
Command Description
:CALCulate2:NULL:OFFSet <n>
:CALCulate2:NULL:STATe <state>
:CALCulate2:NULL:ACQuire
Define null (rel) value (n = rel value).
Enable/disable rel (state = ON or OFF).
Automatically acquire rel value (must
have non-overflowed reading).
Table 8-2
Rel programming example
Command Description
:CALC2:NULL:OFFS 5
:CALC2:NULL:STAT ON
Rel value =5.
Enable rel.