2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-75
Select range
RANGe <n>
:SOURce[1]:CURRent:RANGe <n> Select range for I-Source
:SOURce[1]:VOLTage:RANGe <n> Select range for V-Source
Parameters 2400
<n> = -1.05 to 1.05 Specify I-Source level (amps)
-210 to 210 Specify V-Source level (volts)
DEFault 100µA range (I-Source)
20V range (V-Source)
MINimum 1µA range (I-Source)
200mV range (V-Source)
MAXimum 1A range (I-Source)
200V range (V-Source)
UP Select next higher range
DOWN Select next lower range
2410
<n> = -1.05 to 1.05 Specify I-Source level (amps)
-1100 to 1100 Specify V-Source level (volts)
DEFault 100µA range (I-Source)
20V range (V-Source)
MINimum 1µA range (I-Source)
200mV range (V-Source)
MAXimum 1A range (I-Source)
1100V range (V-Source)
UP Select next higher range
DOWN Select next lower range
2420
<n> = -3.15 to 3.15 Specify I-Source level (amps)
-63 to 63 Specify V-Source level (volts)
DEFault 100µA range (I-Source)
20V range (V-Source)
MINimum 10µA range (I-Source)
200mV range (V-Source)
MAXimum 3A range (I-Source)
63V range (V-Source)
UP Select next higher range
DOWN Select next lower range