2400 Series SourceMeter
®
User’s Manual SCPI Command Reference 18-27
Offset compensated ohms:
*RST
:SENS:FUNC:OFF:ALL
:SENS:FUNC “VOLT”,“CURR”
:SOUR:FUNC VOLT or CURR
if :SOUR:FUNC VOLT then :SOUR:VOLT:STAR <n>;
STOP <n>; MODE SWE
if :SOUR:FUNC CURR then :SOUR:CURR:STAR <n>;
STOP <n>; MODE SWE
:SOUR:SWE:POIN 2
:TRIG:COUN 2
:CALC:MATH:EXPR:NAME “OFFCOMPOHM”
:CALC:STAT ON
:OUTPUT ON
:INIT
:CALC:DATA?
Resistor voltage coefficient
*RST
:SENS:FUNC:ON:ALL
:SENS:RES:MODE MAN
:SOUR:FUNC:ON VOLT or CURR
if :SOUR:FUNC VOLT then :SOUR:VOLT:STAR <n>;
STOP <n>; MODE SWE
if :SOUR:FUNC CURR then :SOUR:CURR:STAR <n>;
STOP <n>; MODE SWE
:SOUR:SWE:POIN 2
:TRIG:COUN 2
:CALC:MATH:EXPR:NAME “VOLTCOEF”
:CALC:STAT ON
:OUTPUT ON
:INIT
:CALC:DATA?