68 CFS/CCS-BA-e-1624
Carefully place fourth adjustment weight into the center of
the weighing plate on display „LoAd 4“.
Wait for stability display, then press .
Ensure that no objects are on the weighing plate on
display „LoAd 0“.
Wait for stability display, then press .
Carefully place fourth adjustment weight into the center of
the weighing plate once more on display „LoAd 4“.
Wait for stability display, then press .
Carefully place third adjustment weight into the center
of the weighing plate once more on display „LoAd 3“.
Wait for stability display, then press .
Carefully place second adjustment weight into the center
of the weighing plate once more on display „LoAd 2“.
Wait for stability display, then press .
Carefully place first adjustment weight into the center of
the weighing plate once more on display „LoAd 1“.
Wait for stability display, then press .
Ensure that no objects are on the weighing plate on
display „LoAd 0“.
Wait for stability display, then press .
After the adjustment the balance will carry out a self-test.
The balance returns automatically into weighing mode.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
( ), repeat adjustment process.
15.2 Model KERN CFS 50K-3
Operation Display
Switch-on balance and during the selftest press .