KFB/KFN-TM-BA_IA-e-1624 19
6.9.2 Non verifiable weighing systems
1. Switch-on balance and during the selftest press .
2. Press subsequently , , the first menu block
„PO CHK“ will be displayed.
3. Press repeatedly until „P3 CAL“ will be displayed.
4. Confirm with ; press repeatedly until „CAL“
appears.
5. Acknowledge using , the current setting is displayed.
Press to confirm; press to select setting.
noLin = adjustment
LineAr = linearization, see chap. 6.10
How to carry out adjustment:
Confirm menu setting „noLin“ by .
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press .
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with .
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press .
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically. An adjusting error or
incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.