14 PFB-BA-e-1630
5.7.2 Adjustment procedure (models PFB-M):
Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
Operation
Start balance by pressing During the
selftest press .
F1 Unt“ is displayed.
Press repeatedly until „tECH“ is
displayed
Press adjustment switch on the lower side of the balance!
Press , „Pin“ is displayed
Enter password:
Press , , subsequently.
„P1 Lin“ is displayed
Press , „P2 CAL“ will be displayed
Press , „UnLoAd“ followed by „LoAd“ is
displayed.
Put the necessary adjustment weight (see
chap.1; Technical Data) on the centre of the
weighing plate.
Adjustment will be carried out automatically after dead
stop control.
Wait for stability display

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PFB Series and is the answer not in the manual?

KERN PFB Series Specifications

General IconGeneral
CalibrationExternal calibration
Weighing Unitsct, g, kg, lb, mo, oz, pcs, t
DisplayLCD with backlight
Power SupplyAC adapter or rechargeable battery
MaterialStainless steel
DimensionsVaries by model
Overall dimensionsVaries by model
Readability0.1 g - 1 g
Readability [d]0.1 g - 1 g

Related product manuals