EasyManua.ls Logo

KERN PFB

KERN PFB
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PFB-BA-e-1322 15
6.7.2 Adjustment procedure (models PFB-M):
Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
Operation
Start balance by pressing During the
selftest press .
F1 Unt“ is displayed.
Press repeatedly until „tECH“ is displayed
Press adjustment switch on the lower side of the balance!
Press , „Pin“ is displayed
Enter password:
Press , , subsequently.
„P1 Lin“ is displayed
Press , „P2 CAL“ will be displayed
Press , „UnLoAd“ followed by „LoAd“ is
displayed.
Put the necessary adjustment weight (see
chap.1; Technical Data) on the centre of the
weighing plate.
Adjustment will be carried out automatically after dead
stop control.
Wait for stability display

Table of Contents

Related product manuals