EasyManua.ls Logo

KERN PFB

KERN PFB
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PFB-BA-e-1322 19
6.8.2 Linearization of non verifiable models (low resolution) and verifiable
models
Balance display
Operation
Start balance by pressing
ON
OFF
. While the balance
carries out a self test, press until „F1 Unt“
appears
Press repeatedly until “tECH“ appears
In case of verifiable models press the adjustment switch on the lower side of the
balance!
Press , „Pin“ appears
Press , and successively, „P1 Lin“
appears
Press anew, „Pin“ appears again
Press successively , and , „LoAd 0“
appears; after the apparition of the stability display
and the signal sound, „LoAd 1“ will be displayed
Put on weight 1 (see table chap.6.8.1)
Wait for the stability display and the signal sound
„LoAd 2“ appears
Remove weight 1 and put on the weight 2
After the stability display and the signal sound „LoAd
3“ appears; remove weight 2 and put on weight 3
After the stability display and the signal sound the
balance carries out a selftest. Remove the weight
during selftest. The balance changes into weighing
mode. Now the linearization is concluded
successfully.

Table of Contents

Related product manuals