PFB-BA-e-1322 19
6.8.2 Linearization of non verifiable models (low resolution) and verifiable
models
Balance display
Operation
Start balance by pressing
ON
OFF
. While the balance
carries out a self test, press until „F1 Unt“
appears
Press repeatedly until “tECH“ appears
In case of verifiable models press the adjustment switch on the lower side of the
balance!
Press , „Pin“ appears
Press , and successively, „P1 Lin“
appears
Press anew, „Pin“ appears again
Press successively , and , „LoAd 0“
appears; after the apparition of the stability display
and the signal sound, „LoAd 1“ will be displayed
Put on weight 1 (see table chap.6.8.1)
Wait for the stability display and the signal sound
„LoAd 2“ appears
Remove weight 1 and put on the weight 2
After the stability display and the signal sound „LoAd
3“ appears; remove weight 2 and put on weight 3
After the stability display and the signal sound the
balance carries out a selftest. Remove the weight
during selftest. The balance changes into weighing
mode. Now the linearization is concluded
successfully.