ALJ/ALS/PLJ/PLS-BA-e-1741 49
Standstill control display
(see chap. 11.6)
Stabil
Stab 1
Standstill control fast very
quiet set-up location
Stab 2
Standstill control fast + exact
quiet set-up location
Stab 3
Standstill control exact very
set-up location
Contrast of the display
(see chap. 11.7)
Contr 1-15
Select contrast
Backlight (see chap. 11.8)
Blt
on Backlight on
off
Backlight off
Auto
Backlight automatically switched
off 3 seconds after achieving
stable weighing value. Changes
in weight or pressing of keys will
automatically result in backlight
switching on again.
AUTO OFF
in stand-by mode
(see chap. 11.9)
time off
disab AUTO-OFF off
2 Min
AUTO-OFF after 2 minutes
without changing the weight
5 Min
AUTO-OFF after 5 minutes
without changing the weight
15 Min
AUTO-OFF after 15 minutes
without changing the weight
Tablet mode tbl mode *tbl off
Not documented
*tbl on
Adjustment
(see chap. 7)
Calib
*E-Cal
Adjustment with external weight,
locked for verifiable units
*Tec Cal
not documented
*Aut Cal
Automatic adjustment with
internal weight.
*I-Cal
Adjustment with internal weight
after having pressed CAL,
locked in verifiable units.
End
* = not available in all models

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PLJ Series and is the answer not in the manual?

KERN PLJ Series Specifications

General IconGeneral
Weighing plate materialstainless steel
Minimum part weight at piece counting0.1g - 2g (depending on model)
Stabilisation time under laboratory conditions2 seconds
Permissible ambient temperature10 °C - 40 °C
InterfaceRS-232
Unitsg, kg, lb, oz