PRS/PRJ/ARS/ARJ-BA-def-0622 133
9.6 Setup for “SET APP CALCULATOR”
To activate the menu see Chapter 7.4 “Activating the two main menus” and Chapter 7.5
“How the menu control operates”.
SETUP
APPLICATION
CALCULATOR SET KEY 1
NAME nnnnn
FACTOR n.nnn e+n
DECIMAL PLACES n
DISPLAY TEXT nnn
PRINTER TEXT nnnnnnnn
SET KEY 2
NAME nnnnn
FACTOR n.nnn e+n
DECIMAL PLACES n
DISPLAY TEXT nnn
PRINTER TEXT nnnnnnnn
SET KEY 3/4
same as for key 1 and 2
When activating the “CALCULATOR” application, each of the four function keys is first
assigned a name, then a particular conversion factor, decimal place definition, display unit
and unit for the printout.
Display in the “CALCULATOR” application:
+ 123.456 g/M
NAME 1 NAME 2 NAME 3 NAME 4
U U U U
Õ” “Ö” “×” “Ø
The value is converted and
printed out.
In programming mode, the previously defined key names are shown above the function
keys.
After pressing a function key, the current value is converted in accordance with the
assigned factor and the result is displayed or printed out after the print key is pressed.
In this way, for example, you can convert the weights of samples of a known size directly
into “grammes per square metre” and these will be displayed.