140 PRS/PRJ/ARS/ARJ-BA-def-0622
9.12 Setup for the statistics program
To activate the menu see Chapter 7.4 “Activating the two main menus” and Chapter 7.5
“How the menu control operates”.
SETUP
STATISTICS
STATISTICS
MODE OFF
MODE STATISTICS
MODE RECORDER
MODE STAT./RECORDER
Statistics program only from
statistics
Only data storing stat. and
storing
NUMBER 100
Number of values which
should be collected
automatically (1 … 999).
MANUAL COLLECTION
COLLECTION ON TIMEBASE
COLLECTION ON LOAD
CHANGE
with function key
“STO”
on timebase
after every change in load
TIMEBASE 2.0
Timebase for “storing” in
seconds
Functions of the statistics program and storing functions:
MODE
- In this function field, you can define whether only the statistics program, only the
storing program or whether both programs are to be used at the same time.
COLLECTION
- With “MANUAL” the function key “STO” must be pressed
for every value which is to be stored.
- With “CHANGE IN LOAD”, the balance automatically stores the
measured value after every change in load.
- With “TIMEBASE” the balance stores the measured value after each defined time
period (factory setting: 2.0 seconds).
TIMEBASE
- Definition of the time period for the collection of data according to “COLLECTION
TIMEBASE” (e.g. every 2 seconds).
NUMBER
- Definition of the number of values after which the store process should be
automatically completed.
NOTE
When storing the first value, a range of ±50 % is defined.
Subsequent values must be within this range, otherwise an error message will appear.