EasyManua.ls Logo

KERN PRS

KERN PRS
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PRS/PRJ/ARS/ARJ-BA-def-0622 149
Command Function
Sn Set stability n n=0 low
n=1 medium
n=2 high
SDTttmm-
jjhhmmss
Set date and time (German) (Tag, Monat, Jahr,
Stunde, Minute, Sekunde)
SDTmmd-
dyyhhmmss
Set Date and Time (English) (Month, Day, Year, Hour, Minutes,
Seconds)
T (ttt) Tare off or set tare to a specific value
Uxnn Set unit x (1...4) of the balance with nn (0=g, 1=mg, 2=kg, ...)
UxS Switch balance to unit x (1...4)
ZERO Zero the balance (provided weight is stable and within the zero-
setting range)
NOTE
Each remote-control command must terminate with “CR” “LF”.
The commands are acknowledged if required.
11.4 Examples for the remote control of the balance
Input Description of the initiated function
D – – – – – Five dashes will be shown
D TEST123. Display will be: tESt123
D The display will be dark
T100 -100.000g (tare set to 100 g)
T1 -1.000 g (tare set to 1g)
T Balance will be tared off
Table 8.5 Examples for remote control

Related product manuals