EasyManua.ls Logo

Keysight Technologies 8163A/B - Page 212

Keysight Technologies 8163A/B
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
212 8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4 Measurement Operations & Settings
command: :TRIGger[n][:CHANnel[m]]:INPut:REARm?
syntax: :TRIGger[n][:CHANnel[m]]:INPut:REARm?
description: Returns the arming response of a channel to an incoming trigger.
parameters: none
response: A boolean value: 0: trigger rearming disabled
1: trigger rearming enabled (default)
example: trig1:inp:rearm? -> 0<END>
affects: All 8163A/B series power meter modules, and 8161x series return loss modules.
dual sensors: Can only be sent to master channel, slave channel parameters are identical.
command: :TRIGger[n][:CHANnel[m]]:OFFSet
syntax: :TRIGger[n][:CHANnel[m]]:OFFSet <value>
description: Sets the number of incoming triggers received before data logging begins.
parameters: <value> - an integer value. (maximum possible value is 1e+9)
response: none
example: trig1:offs 5
affects: All 81636B and 81637B series power meter modules.
command: :TRIGger[n][:CHANnel[m]]:OFFSet?
syntax: :TRIGger[n][:CHANnel[m]]:OFFSet?
description: Returns the number of incoming triggers received before data logging begins.
parameters: none
response: an integer value.
example: trig1:offs?
-> 5<END>
affects: All 81636B and 81637B series power meter modules.