Remote Interface Reference 3
Keysight E364xA User’s and Service Guide 119
State Storage Commands
The power supply has five storage locations in non- volatile memory to store
power supply states. The locations are numbered 1 through 5. You can also assign
a name to each of the locations (1 through 5) for use from the front panel.
*SAV {1|2|3|4|5}
Store (Save) the present state of the power supply to the specified location.
Any state previously stored in the same location is overwritten (no error is
generated).
– A power- on reset (*RST command) does not affect the configurations
stored in memory. Once a state is stored, it remains until it is overwritten or
specifically deleted.
– The state storage feature remembers the states or values of the following
commands:
CURR, CURR:STEP, CURR:TRIG, OUTP, OUTP:REL, TRIG:DEL,
TRIG:SOUR, VOLT, VOLT:STEP, VOLT:TRIG, VOLT:PROT,
VOLT:PROT:STAT, and VOLT:RANG
*RCL {1|2|3|4|5}
Recall the power supply state stored in the specified storage location. When
shipped from the factory, storage locations 1 through 5 are empty.
DISP {OFF|ON} can be stored and recalled in the remote interface mode only.
Going to the local mode automatically sets the display state to ON.