4 Error Messages
146 E364xA User’s and Service Guide
Execution Error Messages
Table 4-1 Execution error messages
-101
Invalid character
An invalid character was found in the command string. You may have inserted a
character such as #, $, or % in the command keyword or within a parameter.
Example: OUTP:STAT #ON
-102
Syntax error
Invalid syntax was found in the command string. You may have inserted a blank
space before or after a colon in the command header, or before a comma.
Example: VOLT:LEV ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used a
comma instead of a colon, semicolon, or blank space, or you may have used a
blank space instead of a comma.
Example: TRIG:SOUR,BUS or APPL 1.0 1.0
-104
Data type error
The wrong parameter type was found in the command string. You may have
specified a number where a string was expected, or vice versa.
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You may have
entered an extra parameter, or you added a parameter to a command that does
not accept a parameter.
Example: APPL? 10

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3640A and is the answer not in the manual?

Keysight Technologies E3640A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3640A
CategoryPower Supply
LanguageEnglish

Related product manuals