EasyManuals Logo

Keysight Technologies E3640A User Manual

Keysight Technologies E3640A
237 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Error Messages 4
E364xA User’s and Service Guide 149
-151
Invalid string data
An invalid character string was received. Check to see if you have enclosed the
character string in single or double quotes.
Example: DISP:TEXT ’ON
-158
String data not allowed
A character string was received but is not allowed for the command. Check the list
of parameters to verify that you have used a valid parameter type.
Example: TRIG:DEL ‘zero’
-160 to -168
Block data errors
The power supply does not accept block data.
-170 to -178
Expression errors
The power supply does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was ignored.
Make sure that the trigger source should be selected to the bus and the trigger
subsystem should be initiated by INIT[:IMM] command.
-213
Init ignored
An INITiate command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a measurement
in progress and place the power supply in the “idle” state.
-221
Settings conflict
Indicates that a legal program data element was parsed but could not be executed
due to the current device state.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example: TRIG:DEL -3
Table 4-1 Execution error messages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3640A and is the answer not in the manual?

Keysight Technologies E3640A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3640A
CategoryPower Supply
LanguageEnglish

Related product manuals