8 Service and Maintenance
222 E364xA User’s and Service Guide
CC source effect (line regulation) E3640A Maximum change: <0.55 mA
E3641A Maximum change: <0.33 mA
E3642A Maximum change: <0.75 mA
E3643A Maximum change: <0.39 mA
E3644A Maximum change: <1.05 mA
E3645A Maximum change: <0.47 mA
CC PARD (Normal mode) all < 4 mA rms
CC PARD (Common mode) all <1.5 µA rms
Table 8-7 CC performance test record