284 Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
18 Save/Recall (Setups, Screens, Data)
Default Setup— recalls the oscilloscope's default setup. This is the same as
pressing the front panel [Default Setup] key. See "Recall the Default
Oscilloscope Setup" on page 32.
Some user settings are not changed when recalling the default setup.
Factory Default— recalls the oscilloscope's factory default settings.
You must confirm the recall because there are no user settings that are left
unchanged.
Performing a Secure Erase
1 Press the [Save/Recall] key.
2 In the Save/Recall Menu, press Default/Erase.
3 In the Default menu, press Secure Erase.
This performs a secure erase of all non-volatile memory in compliance with
National Industrial Security Program Operation Manual (NISPOM) Chapter 8
requirements.
You must confirm the secure erase, and the oscilloscope will reboot when
finished.