174 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Power Supply
Troubleshooting
Figure 8-3 A1 Power Supply Connector at A3 J10
Table 8-3 Power Supply Connector Voltage Levels
Row A B C D E F
1 15VA 15VA 15VA 15VA ACOM +32VA
2 ACOM ACOM ACOM ACOM ACOM ACOM
3 +15VA +15VA +15VA +15VA +15VA +15VA
4 ACOM ACOM ACOM ACOM ACOM ACOM
5 +5.2VA +5.2VA +5.2VA +5.2VA +5.2VA +5.2VA_SENSE
6 +5.2VA +5.2VA +5.2VA +5.2VA +5.2VA +5.2VA
7 ACOM ACOM ACOM ACOM ACOM ACOM
8 ACOM ACOM ACOM ACOM ACOM ACOM
9 +9VA +9VA +9VA +9VA +9VA +9VA_SENSE
10 +9VA +9VA +9VA +9VA +9VA +9VA
11 ACOM ACOM ACOM ACOM ACOM ACOM

Table of Contents