439 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Functional Tests
Leveled Output Power
Troubleshooting Problems with the Output Power Check
Verify that you are using the appropriate power sensor.
Normally, power sensor calibration factors are automatically downloaded to
the power meter when the power meter turns on. If this does not occur,
manually enter the correct calibration factors for the power sensor you are
using.
Verify that the power sensor is properly calibrated to the power meter.