3 Specifications and Characteristics
50 U2000 Series Operating and Service Guide
Figure 3-12 Typical power accuracy at
25 °C for U2000/1/2H models
1,
2
1 Measurement uncertainty 1.9%. At room temperature and excluding power level dependent
terms (zero set, drift, and noise). Refer to Keysight Fundamentals of RF and Microwave Power
Measurements (Part 3) Power Measurement Uncertainty per International Guide (Application
Note 1449-3), 5988-9215EN for more information on measurement uncertainty.
2 After zeroing, 30 minutes of power-on warm-up, and 1024 averages.