3 Specifications and Characteristics
60 U2000 Series Operating and Service Guide
For auto filter, default resolution, and a 10 dB decreasing power step (not
across the switching point), refer to Figure 3- 13.
Figure 3-13 Settling time with auto filter, default resolution, and a 10 dB decreasing power
step (not across the switching point)
-50 dBm
+10 dBm
+2 dBm
-40 dBm
45 ms
90 ms
2.6 s
2.7 s
460 ms
2.8 s
39 s
42 s
42 s
-4 dBm
-10 dBm
-20 dBm
-30 dBm
Normal
Speed
45 ms
82 ms
1.3 s
1.5 s
450 ms
1.6 s
20 s
24 s
24 s
X2
Speed
Maximum sensor
power within a range
U2000/1/2/4A
U2000/1/2H
U2000/1B
+20 dBm
+40 dBm
+12 dBm
+32 dBm
+6 dBm
+26 dBm
0 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
+20 dBm
+10 dBm
0 dBm
-10 dBm
-20 dBm
Minimum sensor
power within a range
Sensor
Dynamic
Range
Ty p ic a l
Settling
Times