Base Matrix Configuration 2
Keysight 34934A User’s Guide 187
Row and Column Wiring
You may connect to the module’s rows and columns by wiring to the screw
terminals marked C1 through C32 and R1 through R16 (H, L) on the silk-screen
(see “34934T-002 Terminal Block: 16x32 Layout” on page 179). The combined
weight of these wires can create significant strain on the wiring; to minimize this
see “Terminal block strain relief” on page 181.
However, you may prefer to use physically separate cabling for the row
connections (e.g. if you connect your DUT to the columns and your measurement
device(s) to the rows). The terminal block’s extension headers J24 and J25 provide
access to the rows; you can make row connections to the central 16 terminals
(pins 3-18) on the supplied terminators you installed in “Placing Safety Interlock
Continuity Jumpers for 16x32” on page 184.
Below is an illustration of the possible locations for making row signal connections
to the 34934T-002: