EasyManua.ls Logo

Keysight DAQ973A - Self-test Procedures; Power-on Self-Test

Keysight DAQ973A
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service and Repair 1
Keysight DAQ970A/DAQ973A Service Guide 23
Self-test Procedures
Power-on self-test
Each time the instrument is powered on, a small set of self-tests are performed.
These tests check that the minimum set of logic and measurement hardware are
functioning properly. Any plug-in modules installed are verified for two-way
communication with the main controller. This self-test is equivalent to the *TST?
SCPI query, and you do not need to remove inputs to the instrument to run the
test. The power-on self-test takes approximately five seconds.
The following procedures shows you how to manually perform a quick instrument
self-test:
Front panel operations
1 Press [Utility] > Self Test.
2 Select Quick Test and view the result: “Self-test passed” or “Self-test failed”.
You can read errors from the error queue by pressing [View] > Errors. Refer to
DAQ970A/DAQ973A Programming Guide for a complete list of self-test errors.
Remote interface operations
1 Connect to the instrument using the remote USB interface.
2 Send *TST? and read the result: pass (+0) or fail (+1).
Use SYSTem:ERRor? to view the errors in the error queue. Refer to DAQ970A/
DAQ973A Programming Guide for a complete list of self-test errors.
All manuals and user guides at all-guides.com

Table of Contents

Related product manuals