Performance Verification 3
Keysight DAQ970A/DAQ973A Service Guide 67
Performance Verification Tests
The performance verification tests are recommended as acceptance tests when 
you first receive the instrument. The acceptance test results should be compared 
against the 90-days test limits. You should use the 24-hours test limits only for 
verification within 24 hours after performing the adjustment procedure. After 
acceptance, you should repeat the performance verification tests at every 
calibration interval.
If the instrument fails performance verification, adjustment or repair is required.
Adjustment is recommended at every calibration interval. If adjustment is not 
made, you must guard band, using no more than 80% of the specifications listed 
in product data sheet, depending on the calibration interval (1 year, 2 year), as the 
verification limits.
Use the Performance Verification Tests to verify the measurement performance of 
the instrument. The performance verification tests use the instrument’s 
specifications listed in product data sheet. You can perform four different levels of 
performance verification tests:
– Self-test: A series of internal verification tests that give a high confidence that 
the instrument is operational.
– Quick verification: A combination of the internal self-tests and selected 
verification tests.
– Performance verification tests: An extensive set of tests that are 
recommended as an acceptance test when you first receive the instrument or 
after performing adjustments.
– Optional verification tests: Tests not performed with every calibration. 
Perform these tests to verify additional specifications or functions of the 
instrument.
NOTE
To ensure optimal test performance:
– Please perform an Autocalibration before you conduct the performance 
verification test.
– Ensure all procedures comply with the recommend test conditions. Please 
refer to “Test Considerations” on page 47 for more information.