EasyManuals Logo

Keysight DAQ973A User Manual

Keysight DAQ973A
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
Calibration Procedures 2
Keysight DAQ970A/DAQ973A Service Guide 37
Input Connections
You will need the DAQM901A multiplexer module to verify or adjust the internal
DMM. The DAQM901A multiplexer module should be connected in slot 300 with
the following connections:
Short (Zero)
Zero test connections for the DAQM901A multiplexer module is best
accomplished using 4-wire copper wires with low-thermal offset measurements.
The 4-wire channels configured for the DAQM901A multiplexer module is channel
n and n+10.
For DC voltage, 2-wire and 4-wire resistance, Channel 303 and 313 are used
for zero test connections as illustrated below:
For DC current, Channel 321 is used for zero test and left OPEN.
All manuals and user guides at all-guides.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ973A and is the answer not in the manual?

Keysight DAQ973A Specifications

General IconGeneral
BrandKeysight
ModelDAQ973A
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals