6 Keysight DAQ970A/DAQ973A Service Guide
WARNING
AC POWER CORD
Removal of the AC power cord is the disconnect method to remove power 
from the instrument. Be sure to allow for adequate access to the power cord 
to permit disconnection from AC power. Use only the Keysight specified 
power cord for the country of use or one with equivalent ratings.
WARNING
SELF-TEST
Before measuring any hazardous voltage or current, run the *TST? query 
from the remote interface, and read the result to verify that the instrument is 
performing properly. The *TST? query is a self-test that returns +0 if the 
instrument passes and +1 if the instrument fails. You can also perform this 
query from the front panel by pressing [Utility] > Self Test > Quick Test. If 
this self-test fails, make sure that the instrument is repaired and passes the 
self-test before continuing.
WARNING
IN CASE OF DAMAGE
Instruments that appear damaged or defective should be made inoperative 
and secured against unintended operation until they can be repaired by 
qualified service personnel.
CAUTION
MEASUREMENT CONSIDERATIONS
This is a sensitive measurement apparatus by design and may have some 
performance loss when exposed to ambient continuous electromagnetic 
phenomenon. Consider using shielded or twisted cables, common mode 
chokes, ferrite clamps, and damping resistors before the input to the 
DAQ970A/DAQ973A.