EasyManuals Logo

Keysight DAQ973A User Manual

Keysight DAQ973A
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
Performance Verification 3
Keysight DAQ970A/DAQ973A Service Guide 75
AC current gain verification test
Input Channel 321
Quick
check
[a]
Error from nominal
Current Frequency Range
AC filter
[b]
24 hours 90 days 1 year 2 years
1 A
1 kHz
1 A >200 Hz
±1.1 mA ±1.3 mA ±1.4 mA ±1.4 mA
5 kHz ±1.1 mA ±1.3 mA ±1.4 mA ±1.4 mA
100 mA
10 Hz
100 mA
>3 Hz ±110 μA ±130 μA ±140 μA ±140 μA
1 kHz
>200 Hz
±110 μA ±130 μA ±140 μA ±140 μA
5 kHz Q ±110 μA ±130 μA ±140 μA ±140 μA
100 μA 1 kHz
10 mA >200 Hz
±4.07 μA ±4.1 μA ±4.1 μA ±4.1 μA
1 mA 1 kHz ±4.7 μA ±4.9 μA ±5 μA ±5 μA
10 mA
1 kHz ±11 μA ±13 μA ±14 μA ±14 μA
5 kHz ±11 μA ±13 μA ±14 μA ±14 μA
1 mA
1 kHz
1 mA >200 Hz
Q ±1.1 μA ±1.3 μA ±1.4 μA ±1.4 μA
5 kHz ±1.1 μA ±1.3 μA ±1.4 μA ±1.4
μA
100 μA
1 kHz
100 μA >200 Hz
±110 nA ±130 nA ±140 nA ±140 nA
5 kHz ±110 nA ±130 nA ±140 nA ±140 nA
[a] Q: Quick performance verification test points.
[b] For Channel 321, select bandwidth AC filter.
All manuals and user guides at all-guides.com