Specifications 1
Keysight E3631A Service Guide 27
[1]
Accuracy specifications are after an 1-hour warm-up and calibration at 25°C.
Voltage <0.01% + 2 mV
Current <0.01% + 250 mA
Programming resolution
+6V output +25V output –25V output
Voltage 0.5 mV 1.5 mV 1.5 mV
Current 0.5 mA 0.1 mA 0.1 mA
Readback resolution
+6V output +25V output –25V output
Voltage 0.5 mV 1.5 mV 1.5 mV
Current 0.5 mA 0.1 mA 0.1 mA
Meter resolution
+6V output +25V output –25V output
Voltage 1 mV 10 mV 10 mV
Current 1 mA 1 mA 1 mA
Transient response time
Less than 50 msec for output recover to within 15 mV following a change in output current from full load to half load or vice
versa
Command processing time
Programming Commands : Maximum time for output to change after receipt of APPLy and SOURce commands : <50 msec
Readback Command : Maximum time to readback output by MEASure? command : <100 msec
The Other Commands : < 50 msec
Tracking accuracy
The ±25V outputs track each other within ± (0.2% of output + 20 mV).