3 Calibration Procedures
84 Keysight E3631A Service Guide
Aborting a calibration in progress
Sometimes it may be necessary to abort a calibration after the procedure has
already been initiated. You can abort a calibration at any time by turning the
power supply off from the front panel. When performing a calibration from the
remote interface, you can abort a calibration by issuing a remote interface device
clear message or by pressing the front-panel "Local" key.