Calibration Procedures 3
Keysight E3631A Service Guide 87
Error messages
The following tables are abbreviated lists of error messages for the E3631A. The
errors listed are the most likely errors to be encountered during calibration and
adjustment. A more complete list of error messages and descriptions is contained
in chapter 5 of the E3631A User's Guide.
System Error Messages
[1]
This error message is only applicable for serial MY53xx6xxx.
Error Error Message
-330 Self-test failed
-350 Too many errors
501 Isolator UART framing error
502 Isolator UART overrun error
503
[1]
SPI data error
511 RS-232 framing error
512 RS-232 overrun error
513 RS-232 parity error
514 Command allowed only with RS-232
521 Input buffer overflow
522 Output buffer overflow
550 Command not allowed in local
800 P25V and N25V coupled by track system
801 P25V and N25V coupled by trigger subsystem