Specifications 1
Keysight E3631A Service Guide 29
AC input ratings (selectable via rear panel selector)
std 115 Vac ± 10%, 47 to 63 Hz
opt 0E3 230 Vac ± 10%, 47 to 63 Hz
opt 0E9 100 Vac ± 10%, 47 to 63 Hz
Maximum input power
350 VA with full load
Cooling
Fan cooled
Operating temperature
0 to 40°C for full rated output. At higher temperatures, the output current is derated linearly to 50% at 55°C maximum
temperature.
Output voltage overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if the output control is set to less than 1 V. If
the output control is set to 1 V or higher, there is no overshoot.
Programming language
SCPI (Standard Commands for Programmable Instruments)
State storage memory
Three (3) user-configurable stored states
Recommended calibration interval
1 year
Dimensions*
212.6 mmW x 132.6 mmH x 348.2 mmD (8.4 x 5.2 x 13.7 in)
*See the next page for detailed information.
Weight
Net 8.2kg (18 lb)
Shopping 11kg (24 lb)