Calibration Procedures 1
Keysight E3632A Service Guide 37
Performance Test Record for E3632A
CV performance test record
Table 1-2 CV performance test record
Test description
Actual
result
Specifications
Upper limit Lower limit
CV programming accuracy @ 0 volts
(DVM reading)
+0.0100 V –0.0100 V
CV readback accuracy @ 0 volts DVM + 0.0050 V DVM – 0.0050 V
CV programming accuracy @ full scale
(DVM reading)
+30.025 V 29.9750 V
CV readback accuracy @ full scale DVM + 0.0200 V DVM – 0.0200 V
CV load regulation
Maximum change: <5
mV
CV line regulation
Maximum change: <5
mV
CV ripple/noise <2 mVp-p, 0.35 mVrms
Load transient response time <50 μsec