Calibration Procedures 1
Keysight E3632A Service Guide 49
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure has
already been initiated. You can abort a calibration at any time by turning the
power supply off from the front panel. When performing a calibration from the
remote interface, you can abort a calibration by issuing a remote interface device
clear message or by pressing the front-panel Local key.