1 Calibration Procedures
38 Keysight E3632A Service Guide
CC performance test record
Table 1-3 CC performance test record
Test description
Actual
resul t
Specifications
Upper limit Lower limit
CC programming accuracy @ 0 A (I
O
)
+0.0100 A –0.0100 A
CC readback accuracy @ 0 A
I
O
+ 0.0050 A I
O
– 0.0050 A
CC programming accuracy @ full scale
(I
O
)
7.0240 A 6.9760 A
CC readback accuracy @ full scale
I
O
+ 0.0155 A I
O
– 0.0155 A
CC load regulation Maximum change: <0.95 mA
CC line regulation Maximum change: <0.95 mA
CC ripple/noise <2 mA
Common mode current noise <1.5 μArms