Keysight EDUX1052G - Figure 4 Setting up Equipment for Internal Trigger Sensitivity Test

Keysight EDUX1052G
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Testing Performance 2
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes Service Guide 33
Figure 4 Setting Up Equipment for Internal Trigger Sensitivity Test
4 Set the output frequency of the signal generator to 10 MHz.
5 Perform these steps to test at the 10 MHz setting:
a Set the signal generator amplitude to about 20 mV
pp
.
b Press the [AutoScale] key.
c Set the time base to 10 ns/div.
d Set channel 1 to 10 mV/div.
e Decrease the amplitude from the signal generator until 0.6 vertical divisions
of the signal (about 6 mV
pp
) is displayed.
50 Ohm Feedthrough
Signal
Generator
Oscilloscope
N Cable