40 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes Service Guide 
3 Calibrating and Adjusting
 
User Calibration
Perform user-calibration:
• Every five years or after 10000 hours of operation.
• If the ambient temperature is >10° C from the calibration temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other 
instruments help determine if you need shorter User Cal intervals.
User Cal performs an internal self-alignment routine to optimize the signal path in 
the oscilloscope. The routine uses internally generated signals to optimize circuits 
that affect channel sensitivity, offset, and trigger parameters. Disconnect all 
inputs and allow the oscilloscope to warm up before performing this procedure.
Performing User Cal will invalidate your Certificate of Calibration. If NIST (National 
Institute of Standards and Technology) traceability is required perform the 
procedures in Chapter 2 in this book using traceable sources.
To perform User Cal
1 Disconnect all inputs from the front panel and allow the oscilloscope to warm 
up before performing this procedure.
2 Make sure user calibration protection is not on by pressing the [Utility] > Options 
> Auxiliary; then, press the Cal Protect softkey to disable calibration protection.
3 Press the [Utility] key; then, press the Service softkey.
4 Begin the User Cal by pressing the Start User Cal softkey.
User Cal Status
Pressing the User Cal Status softkey displays the following summary results of the 
previous User Cal, and the status of probe calibrations for probes that can be 
calibrated. Note that AutoProbes do not need to be calibrated, but InfiniiMax 
probes can be calibrated.
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status: