Keysight N777xC Series Tunable Laser Family Programming Guide 35
Instrument Setup and Status 3
Command: :STATus:OPERation:ENABle
Syntax: :STATus:OPERation:ENABle<wsp><value>
Description: Sets the bits in the Operational Status Enable Summary Mask (OSESM) that enable the contents of the OSESR to affect the
Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR to affect bit 7 of the Status Byte.
Parameters: The bit value for the OSESM as a 16-bit unsigned integer value (0 .. +65535)
The default value is 65535.
Response: none
Example: :stat:oper:enab 128
Command: :STATus:OPERation:ENABle?
Syntax: :STATus:OPERation:ENABle?
Description: Returns the OSESM for the OSESR
Parameters: none
Response: The bit value for the operation enable mask as a 16-bit unsigned integer value (0 .. +65535)
Example: :stat:oper:enab? -> +128
Command: :STATus0:OPERation[:EVENt]?
Syntax: :STATus0:OPERation[:EVENt]?
Description: Returns the Operational Slot Status Event Register (OSSER) of the laser module.
Parameters: none

Table of Contents