Page #36 background imageLoading...
Page #36 background image
36 Keysight N777xC Series Tunable Laser Family Programming Guide
3 Instrument Setup and Status
Response: The results for the individual slot events (a 16-bit unsigned integer value, where 0 value 65535):
Bit
8-16
7
6
5
4
3
2
1
0
Mnemonic
Not used
Not used
Not used
Not used
Slot n: shutter has been opened
Slot n: Zeroing ongoing
Not used
Slot n: Coherence Control has been switched on
Slot n: Laser has been switched on
Decimal Value
256
128
64
32
16
8
4
2
1
Example: :stat0:oper? -> +0
Command: :STATus0:OPERation:CONDition?
Syntax: :STATus0:OPERation:CONDition?
Description: Returns the Operational Slot Status Condition Register of the laser module.
Parameters: none
Response: The results for the individual slot events (a 16-bit unsigned integer value, where 0 value 65535):
Bit
8-16
7
6
5
4
3
2
1
0
Mnemonic
Not used
Not used
Not used
Not used
Shutter open
Zeroing ongoing
Not used
Coherence Control is switched on
Laser is switched on
Decimal Value
256
128
64
32
16
8
4
2
1
Example: :stat0:oper:cond?
-> +0

Table of Contents

Other manuals for Keysight N777 C Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N777 C Series and is the answer not in the manual?

Keysight N777 C Series Specifications

General IconGeneral
BrandKeysight
ModelN777 C Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals