ERTA (Extended Range Transmission Analysis) Mode - Option 209
Configure Measurement Settings
Improve Dynamic Range
The following settings can be made to reduce measurement noise which 
improves Dynamic Range.
Increase source power as much as possible while maintaining a leveled source.
How to set Source Power
—Press Meas Setup 4 
—Then Remote Source
 
—Then Source Power
— Then either Max Power or Leveled - Increase source power until the 
unleveled warning appears. Then decrease power until power is again 
leveled.
To further reduce measurement noise, on the Receiver unit:
— Reduce Attenuation to 5 dB, then turn the Preamp ON. Learn how in 
“Preamplifier Control (Opt 235)” on page 153. 
— Decrease the Res BW. Learn how in “Resolution Bandwidth (Res BW)” on 
page 163.
—Set Frequency Reference to GPS on both units. This requires Option 307 
(Built-in GPS) and two GPS antennae. The GPS reference can be enabled 
for several sweeps. Then if you move to a location without an adequate GPS 
signal, Hold-over Mode will occur. Learn more about GPS as a frequency 
reference in “Frequency Reference Source” on page 492.
Source Enable and Source Mode
When Source Enable [OFF] is selected, or when Source Mode [CW or 
CoupledCW] is selected, then synchronization ends and the two partner units 
revert to general SA (non-tracking) sweeps. However, the units are still in 
partnership, and other settings continue to be synchronized such as 
Frequencies, ResBw, and Points.