404                                                                                                                     Keysight N9927-90001 User’s Guide
Noise Figure Mode (Option 356 – CPU2 Only)
How to Define a Noise Figure Measurement
— Gain Specify Gain or 
— Noise Temperature
 Specify Noise Temperature or
— YFactor
 Specify Y-Factor
b. Then modify the settings based on the selected noise figure type
2. Select Measure Setup
—choose:
— Noise Source
 or “DUT or” on page 406, “Integration or” on 
page 407, or “Uncertainty
” on page 407.
Opens a menu of softkeys to define the Noise Figure’s ENR Mode 
Spot or Table and uncertainty values.
Then choose:
— ENR Mode Spot Table
 You can enter ENR data for the 
noise source you are using as a table of values or as a 
single spot value. The values held in the table can be used 
for measurements at a range of frequencies as well as at a 
fixed frequency. 
The single spot value is used either for measurements at a 
single frequency, or for measurements across a range of 
frequencies that is narrow enough such that the ENR value 
does not change significantly across that range. (Default is 
Table).
— ENR Spot Values
 Opens a menu to enter a spot value 
either for measurements at a single frequency or for 
measurements across a range of frequencies that is narrow 
enough such that the ENR value does not change 
significantly across that range (Inactive with ENR Mode set 
to Table).
Then choose:
——ENR
 sets the ENR spot value in dB (default is 
15 dB).
——ENR Uncert.
 sets the ENR spot uncertainty value 
in dB (default is 0.000 dB).
——On | 
Γ| sets the On value (hot) of the ENR spot 
value’s 50
Ω match, Γ (0.000 to 1.000).
——Off | 
Γ| sets the Off value (cold) of the ENR spot 
value’s 50
Ω match, Γ (0.000 to 1.000).
——ENR Uncert. Cov. [2
σ] Opens a menu to set the 
value of the ENR uncertainty coverage.
Then choose: 1
σ, 2σ (default), or 3σ.