EasyManua.ls Logo

Keysight PXA N9030A - Page 304

Keysight PXA N9030A
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 System Functions
System
Notes The values displayed on the screen are only updated upon entry to the screen and not updated while
the screen is being displayed.
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:SYSTem:PON:TIME?
Example :SYST:PON:TIME?
Notes Value is the time since the most recent start-up in seconds.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:CURRent?
Example :CAL:TEMP:CURR?
Notes Value is in degrees Centigrade.
Value is invalid if using default alignment data (Align Now, All required)
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TIME:LALL?
Example :CAL:TIME:LALL?
Notes Value is the elapsed time, in seconds, since the last successful Align Now, All or Align Now, All but
RF was executed.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:LALL?
Example :CAL:TEMP:LALL?