8 Frequency Scan Measurement
AMPTD Y Scale
Key Path
AMPTD Y Scale
Remote Command
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Example POW:ATT 10
POW:ATT?
Preset 10 dB
State Saved Saved in instrument state.
Min 0 dB
Max 50 dB (CXA)
60 dB (EXA)
70 dB (MXA, PXA& N9038A)
Default Unit dB
Backwards
Compatibility SCPI
:INPut[1]|2:ATTenuation
Backwards Compatibility
SCPI Notes
This command above is included for ESU compatibility. The suffix [1]|2 is irrelevant to the
measurement, they are added solely for remote language compatibility. This SCPI is Meas Local and
Context Sensitive.
Initial S/W Revision A.07.00
Scale/ Div
Sets the units per division of the vertical scale in the logarithmic display.
Key Path
AMPTD Y Scale
Remote Command
:DISPlay:FSCan:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <rel_ampl>
:DISPlay:FSCan:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:PDIVision?
Example DISP:FSC:VIEW:WIND:TRAC:Y:PDIV 5 dB
DISP:FSC:VIEW:WIND:TRAC:Y:PDIV?
Preset 10.00 dB
State Saved Saved in instrument state.
Min 0.10 dB
Max 20 dB
Default Unit dB
Initial S/W Revision A.07.00