Loading...
6 System Functions
System
List installed Options (Remote Command Only)
Lists the installed options that pertain to the instrument (signal analyzer). .
Mode All
Remote Command
:SYSTem:OPTions?
Example :SYST:OPT?
Notes The return string is a comma separated list of the installed options. For example:
503,P03,PFR”
:SYSTem:OPTions? and *OPT? are the same.
State Saved No
Initial S/W Revision Prior to A.02.00
Lock the Front-panel keys (Remote Command Only)
Disables the instrument keyboard to prevent local input when the instrument is controlled remotely.
Annunciation showing a K” for ‘Klock (keyboard lock) alerts the local user that the keyboard is locked.
Klock is similar to the GPIB Local Lockout function; namely that no front-panel keys are active with the
exception of the Power Standby key. (The instrument is allowed to be turned-off if Klock is ON.) The Klock
command is used in remote control situations where Local Lockout cannot be used.
Although primary intent of Klock is to lock-out the front panel, it will lock-out externally connected
keyboards through USB. Klock has no effect on externally connected pointing devices (mice).
The front panel ‘Local key (Cancel/Esc) has no effect if Klock is ON.
Mode All
Remote Command
:SYSTem:KLOCk OFF|ON|0|1
:SYSTem:KLOCk?
Example :SYST:KLOC ON
Notes Keyboard lock remains in effect until turned-off or the instrument is power-cycled
Preset Initialized to OFF at startup, unaffected by Preset
State Saved No
Initial S/W Revision Prior to A.02.00
List SCPI Commands (Remote Command Only)
Outputs a list of the valid SCPI commands for the currently selected Mode.
Remote Command
:SYSTem:HELP:HEADers?
Example :SYST:HELP:HEAD?
Notes The output is an IEEE Block format with each command separated with the New-Line character (hex