8 Frequency Scan Measurement
Marker
Example CALC:FSC:MARK:TRAC 1
CALC:FSC:MARK:TRAC?
Notes A marker may be placed on a blanked and/or inactive trace, even though the trace is not visible
and/or updating.
Couplings Sending the remote command causes the addressed marker to become selected.
Preset 1
State Saved Saved in instrument state.
Range 1|2|3
Readback Text Trace N where N is the trace number to which the marker is currently assigned.
Initial S/W Revision A.07.00
Couple Markers
When this function is on, moving any marker causes an equal X Axis movement of every other marker that
is not off. By equal X Axis movement” we mean that we preserve the difference between each marker’s X
Axis value (in the fundamental x-axis units of the trace that marker is on) and the X Axis value of the marker
being moved (in the same fundamental x-axis units).
This may result in markers going off screen.
Key Path
Marker
Remote Command
:CALCulate:FSCan:MARKer:COUPle[:STATe] OFF|ON|0|1
:CALCulate:FSCan:MARKer:COUPle[:STATe]?
Example CALC:FSC:MARK:COUP ON Sets Couple Markers on.
Preset Off, presets on Mode Preset and All Markers Off
State Saved Saved in instrument state.
Initial S/W Revision A.07.00
All Markers Off
Turn all markers off.
Key Path
Marker
Remote Command
:CALCulate:FSCan:MARKer:AOFF
Example CALC:FSC:MARK:AOFF
Initial S/W Revision A.07.00