6 System Functions
System
alignment that pertains to the instrument must be transferred to the resident hard drive after a CPU or
hard drive is replaced. This utility facilitates backing-up and restoring the alignment data.
This utility allows the operator to navigate to any location of the Windows file system. It is intended that
the operator use a USB memory device or Mapped Network Drive to back up the alignment data to storage
outside of the instrument.
Key Path
System, Alignments
Initial S/W Revision A.02.00
Key Path
System, Alignments
Mode All
Remote Command
:CALibration:DATA:DEFault
Example :CAL:DATA:DEF
Couplings Sets Auto Align to Off. Sets bit 14 in the Status Questionable Calibration register. The Error Condition
message Align Now, All required” is generated.
Initial S/W Revision Prior to A.02.00
Alignment Data Wizard
The Backup or Restore Alignment Data wizard guides you through the operation of backing-up or restoring
the alignment data.
The following dialogue boxes operates without a mouse or external keyboard when you use the default file
names.
EMI Receiver Mode Reference 309