EasyManua.ls Logo

Keysight PXA N9030A - Query the Mechanical Relay Cycle Count

Keysight PXA N9030A
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 System Functions
System
Query the Mechanical Relay Cycle Count
Return the count of mechanical relay cycles. For N9038A model, there are additional 2 Mechanical Relays
which are <N9038A Input2>, <N9038A Bypass>.
Remote Command
:SYSTem:MRELay:COUNt?
Example :SYST:MREL:COUN?
Notes Query Only
The return value is a comma separated list of the individual counts for each mechanical relay.
The position of the relays in the list is:
<Cal Signal>,<AC/DC>,<2dB #1 Atten>,<2dB #2 Atten>,<6dB Atten>,<10dB Atten>,<20dB
Atten>,<30dB Atten>,<Fixed Atten>,<Low Noise Path Switch>,<Presel Bypass>,<N9038A Input2>,
<N9038A Bypass>
Items in the list not pertaining to your particular hardware configuration will return as 999 for those
items.
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.08.00
Query the Operating Temperature Extremes
Returns the low operating temperature extreme value. The value survives a power-cycle and is the
temperature extreme encountered since the value was reset by the factory or service center.
Mode All
Remote Command
:SYSTem:TEMPerature:LEXTreme?
Example :SYST:TEMP:LEXT?
Notes Value is in degrees Celsius at which the lowest operating temperature has been recorded since 1st
power-up.
State Saved No
Initial S/W Revision Prior to A.02.00
Mode All
Remote Command
:SYSTem:TEMPerature:HEXTreme?
Example :SYST:TEMP:HEXT?
Notes Value is in degrees Celsius at which the highest operating temperature has been recorded since 1st
power-up.
State Saved No
Initial S/W Revision Prior to A.02.00
Query the Elapsed Time since 1
st
power on
Returns the elapsed on-time in minutes since 1st power-on.
EMI Receiver Mode Reference 351